# FastGene qFYR (Plus) Real-Time PCR System

---
url: https://www.nippongenetics.eu/en/products/lab-instruments/qpcr-thermal-cycler/fastgene-qfyr-plus-real-time-pcr-plus-system/
language: en
type: product
modified: 2026-04-29T09:06:33+02:00
---

## Real-time fluorescent quantitative PCR system |  4+1 or 6 channel system | 96-wells

## Product Details

- **SKU**: FG-QPTC01, FG-QPTC02
- **Stock status**: outofstock
- **Categories**: Lab Instruments, qPCR Thermal Cycler

- Fast detection − Probe &amp; dye-based assays
- 4+1 or 6 different colour channels - perfect for multiplex qPCR **[(view channels)](#channels)**
    - 4+1 channels: 4 different colour channels + 1 additional second FAM/SYBR channel for fast dual mode melt curve analysis
    - 6 channels
- Standard plate format − 96-well block with gradient function
- Highest sensitivity
- Outstanding homogeneity across all wells of a 96-well plate
- Software included − Easy-to-use and intuitive
- Integrated Precision Melt Analysis Tool (for probe-based allelic discrimination)
- Many applications − Gene expression analysis, genotyping, HRM (high resolution melt), pathogen detection, genetic analysis

[PRODUCT INQUIRY](#section-inquiry_form)

### Details

### Introduction

##### Highest precision instrument

The FastGene® qFYR is a highest precision instrument for performing quantitative polymerase chain reaction (qPCR) experiments. qPCR is a well-established method for the sensitive detection and quantification of nucleic acids. During a measurement, the target DNA or RNA sequence is amplified, while a cycle-dependent increase of a fluorescent signal is detected in real-time.

##### The device for multiple applications

The FastGene® qFYR was developed to meet highest laboratory standards and deliver reliable performances for various real-time qPCR applications:

- Gene expression analysis
- Absolute and relative quantification
- Endpoint qualitative analysis
- Genotyping
- Gene mutation analysis
- Pathogen detection
- GMO detection
- Protein stability screening
- miRNA analysis
- Melting curve analysis

### Innovative Technology

##### High sensitivity optical design

The FastGene® qFYR has a unique, patented optical detection system. It combines a high quality MPPC detector with a Fresnel lens that has a short focal length. The resulting short distance from detector to sample reduces signal loss and cross-talk between samples, and generally improves signal sensitivity.

![FG qFYR Real Time PCR System | High optical detection | Drawing](https://www.nippongenetics.eu/app/uploads/fg-qfyr-real-time-pcr-system-high-optical-detection-300x156.png)

##### Superior thermal cycler performance

The FastGene® qFYR delivers excellent temperature control, temperature precision (± 0.2 °C) and uniformity over the entire 96-well plate. The unique hollowed out thermal block design reduces its overall weight and ensures rapid heating and cooling rates (up to 6 °C per second) for fast qPCR protocols. The system also uses state-of-the-art Peltier components for highest reaction quality and performance stability, providing reliable and consistent qPCR results.

![](https://www.nippongenetics.eu/app/uploads/fg-qfyr-thermoblock.png)
<a name="channels"></a>

##### Fast measurement with high performance LED channels

Due to the small distance between the scanning head and the plate, the FastGene® qFYR has an outstanding level of sensitivity and virtually no cross talk between the individual wells.
4 or 6 different colour channels are installed in the scanning head to be able to excite all commonly used qPCR dyes in a single plate scan (8 sec).

The FastGene® qFYR is available in 2 different versions, with:

- 4+1 channels (FastGene® qFYR: FG-QPTC01) and
- 6 channels (FastGene® qFYR **Plus**: FG-QPTC02).

**I) FastGene® qFYR (FG-QPTC01)**

The FastGene® qFYR (FG-QPTC01) has 4+1 channels: 4 different colour channels and 1 second FAM channel.
The dual FAM/SYBR channel allows melt curve and high resolution melt experiments to be performed in fast dual mode, halving the measuring time in this channel.

Channel 1: FAM/SYBR Green etc.
Channel 2: VIC/JOE/HEX/TET etc.
Channel 3: ROX/Texas Red etc.
Channel 4: Cy5 etc.
Channel 5: FAM/SYBR Green etc.

**II) FastGene® qFYR Plus (FG-QPTC02)**

The FastGene® qFYR Plus (FG-QPTC02) has 6 different colour channels.

Channel 1: FAM/SYBR Green etc.
Channel 2: VIC/JOE/HEX/TET etc.
Channel 3: ROX/Texas Red etc.
Channel 4: Cy5 etc.
Channel 5: Cy5.5/Quasar 705/Alexa Fluor 680
Channel 6: ATTO 42

### Software | Video tutorials

##### The powerful user software has it all

The analysis software (FastGene® qFYR Analysis Studio) impresses with its particularly simple and operation. The menu is clearly structured and intuitively arranged. It responds to the needs of the user for different experimental setups, and personalized settings can be easily adjusted. Integrated analysis algorithms allow many steps, such as baseline subtraction or Cq value threshold calculation, to be performed automatically.
Absolute and relative quantification of nucleic acids can also be automated. The software allows you to save predefined analysis settings for auto-exporting run data into a format of choice, including Excel, PDF, txt export format.

- Clearly structured and user-friendly
- Intuitive navigation
- Customizable settings
- Automatic analysis with integrated algorithms
- Modules for multiple applications
- Modern UI design

![](https://www.nippongenetics.eu/app/uploads/qfyr-software-setup.png)

##### Video tutorial – Complete webinar | 25:22 min

See how easy it is to (1) **install the software**, (2) **set up the experiments** and (3) **analyse the data**.

 Um das YouTube-Video zu sehen, müssen Sie Cookies akzeptieren.

  Cookie-Einstellungen







 ![](https://www.nippongenetics.eu/app/uploads/qfyr-software-tutorial-video-thumbnail-1024x576.png) ![]()



##### Short video: How to set up your Real-Time PCR experiment in under 3 minutes | 2:51 min

 Um das YouTube-Video zu sehen, müssen Sie Cookies akzeptieren.

  Cookie-Einstellungen







 ![FasteGene qFYR Real-Time PCR System | Experiment setup | Video thumbnail](https://www.nippongenetics.eu/app/uploads/qfyr-software-setup-video-thumbnail-1024x576.png) ![]()



##### Short video: How to analyse and export your real-time data in under 2 minutes | 1:29 min

 Um das YouTube-Video zu sehen, müssen Sie Cookies akzeptieren.

  Cookie-Einstellungen







 ![FasteGene qFYR Real-Time PCR System | Data analysis and export | Video thumbnail](https://www.nippongenetics.eu/app/uploads/qfyr-software-analyse-video-thumbnail-1024x576.png) ![]()



### Application data

##### Generate high quality data

The combination of innovative optics and a high-precision thermal block ensure ideal amplification conditions. This makes the FastGene® qFYR ideal for various quantitative real-time PCR applications.

##### Outstanding homogeneity

FastGene® qFYR shows great homogeneity across all wells of a 96-well plate.

![](https://www.nippongenetics.eu/app/uploads/fg-qfyr-consisten-amplification-data-300x179.png)

_Figure 1: Amplification of 1 ng plasmid DNA in all 96 wells of a PCR plate. The mean Cq value at a cycle number of 13.89 +/- 0.055 was determined automatically, illustrating highly homogenous amplification results obtained by the FastGene® qFYR._

##### Powerful multiplexing

FastGene® qFYR Plus can discriminate up to 6 different targets in a single reaction well, helping you save time, conserve reagents, and reduce samples usage without sacrificing precision.

![](https://www.nippongenetics.eu/app/uploads/qfyr-multiplex-amplification-plot-300x190.png)

_Figure 2: Multiplex amplification of six mammalian housekeeping genes (GAPDH, B2M, ACTG1, PGK1, HPRT1, and ACTB ). Each gene was detected using a different fluorescent probe: FAM – GAPDH, HEX – B2M, TexasRed – ACTG1, Cy5.5 – PGK1, ATTO425 – HPRT1, Cy5 – ACTB._

##### Accurate quantification

The broad dynamic range of the FastGene® qFYR ensures a reliable and accurate quantification.

![](https://www.nippongenetics.eu/app/uploads/fastgene-qfyr-application-data-accurate-amplification.png)

_Figure 3: The amplification plot shows the log of change in normalized reporter fluorescence against the cycle number. The amplification took place from plasmid DNA with AMP specific primers in a 10-fold dilution series ranging from 1 ng to 1×10-6 ng plasmid DNA. The generated standard curve shows 100 % efficiency._

##### 1.3-fold target discrimination

With the FastGene® qFYR, smaller concentration differences can be distinguished with high accuracy, underlining the high sensitivity of the device.

![](https://www.nippongenetics.eu/app/uploads/fastgene-qfyr-application-data-high-accuracy-300x171.png)

_Figure 4: The amplification of plasmid DNA was carried out using AMP-specific primers with a 2-fold dilution series starting at 0.1 ng with additional dilutions of 1:1.3, 1:1.4, 1:1.5 and 1:1.6. Concentration differences can be detected up to 1:1.3-fold dilution._

##### Melt curve analysis

The Precision Melt Analysis Tool is a powerful method for probe-based allelic discrimination. This function is already included in the FastGene® qFYR analysis software and does not need to purchased additionally. ![](https://www.nippongenetics.eu/app/uploads/fastgene-qfyr-application-data-melt-curve-analysis-300x179.png)

_Figure 5: The difference plot of the high-resolution melting curve allows the discrimination between SNPs extracted from blood samples for brown eyes and two variants of blue eyes._

##### Gradient qPCR for optimal annealing

The thermal gradient function of the FastGene® qFYR can be used to determine the optimal annealing temperature of a specific target.

![](https://www.nippongenetics.eu/app/uploads/fastgene-qfyr-application-data-melt-curve-analysis-1.png)

_Figure 6: Plasmid amplification with IC green dye was carried out in a 10-fold dilution series from 1 ng to 1×10-6 ng. This example of a thermal gradient experiment with 55 °C and 68 °C shows 55 °C to be the optimal annealing temperature with a 100% efficiency._

### Specifications

##### Thermal cycler

**FastGene® qFYR** (FG-QPTC01)
4+1 channel system**FastGene® qFYR Plus** (FG-QPTC02)
6 channel system**![FastGene qFYR Real Time qPCR Cycler - front view](https://www.nippongenetics.eu/app/uploads/qFYR-real-time-pcr-system-front-view-150x150.png)****![FastGene qFYR Plus Real Time qPCR Cycler - front view](https://www.nippongenetics.eu/app/uploads/fg-qfyr-plus-real-time-pcr-system-fg-qptc02-front-view-150x150.png)**Block capacity9696Sample volume5–100 µL5–100 µLHeating and cooling methodPeltier (6 temperature control modules)Peltier (6 temperature control modules)Temperature control technologyHollow-out module combined with edge temperature compensation technologyHollow-out module combined with edge temperature compensation technologyThe maximum heating and cooling rate of the block6.0 °C/s6.0 °C/sThe average heating and cooling rate of the sample4.0 °C/s4.0 °C/sTemperature range4–100 °C4–100 °CTemperature accuracy± 0.2 °C± 0.2 °CTemperature uniformity± 0.2 °C± 0.2 °CTemperature setting range4–100 °C4–100 °CHeated lidElectronic automatic lidElectronic automatic lidGradient zones12 columns12 columnsTemperature gradient temperature range1–36 °C1–**40 °C**Linear dynamic range10 orders of magnitude: 1-1010 copies10 orders of magnitude: 1-1010 copiesSoftware nameFastGene®qFYR Analysis StudioFastGene®qFYR Analysis Studio##### Optical detection

**FastGene® qFYR** (FG-QPTC01)
4+1 channel system**FastGene® qFYR Plus** (FG-QPTC02)
6 channel systemExcitation light source**4** single-color high-efficiency LEDs
(maintenance-free, working life &gt;100,000 hours)**6** single-color high-efficiency LEDs
(maintenance-free, working life &gt;100,000 hours)Detector deviceHighly sensitive MPPC (SiPM – solid-state photomultiplier)Highly sensitive MPPC (SiPM – solid-state photomultiplier)Scanning principleTime-resolved real-time scanTime-resolved real-time scanFluorescent channel number4 channels (2x FAM)**6 channels**Application to fluorescent- Channel 1: FAM/SYBR Green etc.
- Channel 2: VIC/JOE/HEX/TET etc.
- Channel 3: ROX/Texas Red etc.
- Channel 4: Cy5 etc.
- Channel 5: FAM/SYBR Green etc.

- Channel 1: FAM/SYBR Green etc.
- Channel 2: VIC/JOE/HEX/TET etc.
- Channel 3: ROX/Texas Red etc.
- Channel 4: Cy5 etc.
- **Channel 5: Cy5.5/Quasar 705/Alexa Fluor 680**
- **Channel 6: ATTO 425**

Dye compatibility- FAM/SYBR Green
- VIC/HEX/TET/JOE
- ROX/Texas Red, Mustang Purple
- Cy5/LIZ

- FAM/SYBR Green
- VIC/HEX/TET/JOE
- ROX/Texas Red, Mustang Purple
- Cy5/LIZ
- **Cy5.5/Quasar 705/Alexa Fluor 680**
- **ATTO 425**

Detector positionTop of the blockTop of the blockTop of the blockDetection sensitivity1 copy of the target sequence1 copy of the target sequenceSystem sensitivityDistinguishable 1.33-fold copy number difference in singleplex reactionsDistinguishable 1.33-fold copy number difference in singleplex reactionsDetection time- Standard mode (full channel): 8.5 seconds/96-well plate
- Fast mode (dual FAM): 4 seconds/96-well plate

- Standard mode (full channel): 8.5 seconds/96-well plate

Exitation/detection range- Excitation range: 455–650 nm;
- Scope of test: 510–715 nm

- Excitation range: **415–685 nm**;
- Scope of test: **455–745 nm**

##### Physical specifications

**FastGene® qFYR** (FG-QPTC01)
4+1 channel system**FastGene® qFYR Plus** (FG-QPTC02)
6 channel systemInstrument specifications- Dimensions: 35 cm x 52 cm x 37 cm (length x wide x height)
- Weight: 25 kg

- Dimensions: 35 cm x 52 cm x 37 cm (length x wide x height)
- Weight: 25 kg

Environmental conditions- Temperature: 10–30 °C
- Humidity: ≤85 % RH no condensation
- Atmosheric pressure: 85.0 kPa ~ 106.0 kPa
- Altitude: &lt; 2000 m

- Temperature: 10–30 °C
- Humidity: ≤85 % RH no condensation
- Atmosheric pressure: 85.0 kPa ~ 106.0 kPa
- Altitude: &lt; 2000 m

Power supply specification- Power supply voltage: AC 100–240 V
- Frequency: 50 Hz
- Power: 750 W

- Power supply voltage: AC 100–240 V
- Frequency: 50 Hz
- Power: 750 W

Data communication interfaceUSB portsUSB ports### Compatibility &amp; Consumables

##### Get the right consumables for FastGene® qFYR

The FastGene® qFYR is compatible with **low-profile (0.1 mL)** PCR tubes/8-well PCR tube strips with transparent, **flat tops** as well as **non-skirted or half-skirted low profile** 96 well PCR reaction plates ([See product category: PCR Plastic (link)](https://www.nippongenetics.eu/en/product-category/laboratory-plastic/pcr-plastic/)):

- **Single tubes:** Low-profile (0.1 mL) clear/white PCR thin-walled single tube, flat-topped transparent optical tube cover, e.g.:
    - PCR Tubes 0.1 mL with flat caps [(FG-011F)](https://www.nippongenetics.eu/en/products/laboratory-plastic/pcr-plastic/single-tubes-2/pcr-tubes-0-1-ml-with-flat-caps-1000/)
- **8-well strips:** Low-profile (0.1 mL) clear/white PCR thin-walled 8-tube, flat-top transparent optical tube cover, e.g.:
    - 8-well PCR Tube Strips 0.1 mL with flat cap strips[ (FG-017FC)](https://www.nippongenetics.eu/en/products/laboratory-plastic/pcr-plastic/8-well-strips/8-well-pcr-strips-0-1-ml-and-flat-cap-strips-120/)
    - 8-well PCR Tube Strips 0.1 mL with flat caps[ (FG-018WF)](https://www.nippongenetics.eu/en/products/laboratory-plastic/pcr-plastic/8-well-strips/8-well-pcr-strips-0-1-ml-with-flat-caps-120/)
    - 8-well White PCR Tube Strips 0.1 mL + optical Flat Cap Strips [(FG-019FC)](https://www.nippongenetics.eu/en/products/laboratory-plastic/pcr-plastic/8-well-strips/8-well-white-pcr-tube-strips-0-1ml-optical-flat-cap-strips-125/)
- **96-well plates:** Low-profile (0.1 mL) clear/white PCR thin-walled 96- well plate with no hemline or hemline, e.g.:
    - FastGene® White 96-well PCR Plate 0.1 mL, semi-skirted [(FG-2102)](https://www.nippongenetics.eu/en/products/laboratory-plastic/pcr-plastic/96-well-plates/fastgene-white-96-well-pcr-plate-0-1-ml-semi-skirted/)
    - FastGene® 96-well PCR Plate 0.1 mL, non-skirted[ (FG-170350)](https://www.nippongenetics.eu/en/products/laboratory-plastic/pcr-plastic/96-well-plates/fastgene-96-well-pcr-plate-0-1-ml-non-skirted/)
    - FastGene® 96-well Fast PCR Plate 0.1 mL [(FG-03890-50)](https://www.nippongenetics.eu/en/products/laboratory-plastic/pcr-plastic/96-well-plates/96-well-fast-pcr-plate-0-1-ml-25/)

**! It is not compatibel with high-profile (0.2 mL) PCR reaction tube and convex tube covers !**

**_Info: If the sample tubes are just a few, use 8-well tubes for support on both sides, or use empty single tubes in the four corners to prevent uneven force on the plate slot._**

##### Dye compatibility

Our FastGene® IC Green and Probe Mixes are the ideal reagents for the FastGene® qFYR.
Find the right dye mix for your application:[ PCR reagents and enzymes – Quantitative PCR (link)](https://www.nippongenetics.eu/en/product-category/pcr-reagents-and-enzymes/qpcr/).

F1: FAM/SYBR Green
F2: VIC/HEX/TET/JOE, JUN
F3: ROX/Texas Red, Mustang Purple
F4: Cy5/LIZ
F5: Cy5.5/Quasar 705/Alexa Fluor 680 (FastGene® qFYR **Plus** only)
F6: ATTO 425 (FastGene® qFYR **Plus** only)

##### Computer configuration requirements

The computer installed with the FastGene® qFYR Analysis Studio must have the following configuration:
**Processor**: Intel or AMD dual-core CPU,2.8GHz
**Memory**: 4 GB
**Hard disk**: 500 GB
**Network card**: 10 M/100M adaptive network card (optional)
**Display resolution**: 1440×900 or higher
**Operating system**: Windows 10 or above &amp; Office Word/Excel 2007 or above.

For furher details, please refer to the user manual, section [‘Installation Requirements’](https://www.nippongenetics.eu/app/uploads/manual-fastgene-qfyr-real-time-pcr-system.pdf).

### Downloads &amp; Service

##### Service for customers

##### [REgister your instrument](https://www.nippongenetics.eu/en/instrument-registration/)

###### Activate our exclusive service now &amp; enjoy your benefits

- **Priority access** to latest software updates
- **Fast line** to our expert service team
- **Sneak peak** at new features and accessories

[More information](https://www.nippongenetics.eu/en/instrument-registration/)

##### Manual

[Manual – FastGene qFYR Real-Time PCR System (FG-QPTC01, FG-QPTC02)](https://www.nippongenetics.eu/app/uploads/manual-fastgene-qfyr-real-time-pcr-system.pdf)

##### Brochure

[Brochure – FastGene qFYR Real-Time PCR System (FG-QPTC01. FG-QPTC02)](https://www.nippongenetics.eu/app/uploads/brochure-fastgene-qfyr-real-time-pcr-system.pdf)

##### Whitepaper

[Whitepaper | How a powerful multiplex PCR System simplifies laboratory workflows | FastGene qFYR (Plus)](https://www.nippongenetics.eu/app/uploads/whitepaper-how-a-powerful-multiplex-pcr-system-simplifies-laboratory-workflows.pdf)

##### Technical Note

[Technical Note | FastGene qFYR | ROX reaction mix compatibility](https://www.nippongenetics.eu/app/uploads/tcn-qptc01-04-fastgene-qfyr-rox-reaction-mix-compatibility.pdf)

 <a id="product-reviews"></a>###  Reviews

There are no reviews yet.



Only logged in customers may leave a review.