Details
The compact specialist microvolume photometer

The FastGene® NanoView Photometer is a small micro sample analyzer with a very easy and convenient user interface.
Nucleic acids and proteins are quantified with highest reliablility via absorption measurement at 260 nm, 280 nm and 600 nm.
Technology
The FastGene® NanoView Photometer uses a special light path optimization technique (slope algorithm) to enable measurement of a wide range of absorbance.
Measurement and configuration

The stated factor of each pre-configurated measurement mode is set as the default value. Therefore, no adjustment is needed.
- Nucleic acids: dsDNA | ssDNA | RNA | Factor=?*
- Proteins: OD600 | Protein | Lysozyme | BSA | IgG | Factor=?*
*Maximal flexibility and specificity is achieved by the input mode (Factor=?) which enables the user to set a custom-specific factor value which is different from the default value.
Video
Introduction video FastGene® NanoSpec and FastGene® NanoView Photometers
Usability (free demo)
Measurement modes and configuration

The menu of the FastGene® NanoView is very simple and intuitive. 10 pre-configurated measuring modes ensure secure utilization and highest reliability.
Maximal flexibility and specificity is achieved by the input mode (Factor=?) which enables the user to set a custom-specific factor value which is different from the default value.
Main menu:

Free demonstration
We would be happy to demonstrate the ease of use of the FastGene® NanoView photometer.
Just ask for a free product demonstration (online or on site).
Product comparison

The FastGene® photometers NanoView (FG-NP01) and NanopSpec (FG-NP02) are microvolume spectrophotometers for easy and reliable biolmolecule (nucleic acids and proteins) analysis. Both photometers include a microvolume reader and cuvette holder, glove-compatible touchscreen and convenient user interface.
FastGene® NanoSpec: Full spectrum Xenon flash lamp: 190-850 nm (microvolume and cuvette reader)
FastGene® NanoView: Three different wavelength LEDs: 260 nm, 280 nm (microvolume reader), 600 nm (cuvette reader)
Detailed comparison
|

|

|
Specification
|
FastGene® NanoSpec Photometer (link)
(FG-NP01)
|
FastGene® NanoView Photometer
(FG-NP02)
|
General specifications |
|
|
Photometric system |
Microvolume spectrophotometer |
Microvolume photometer |
Applications |
Nucleic acid analysis (dsDNA, ssDNA, RNA), colorimetric protein analysis (BCA, Bradford, Lowry), cell density (OD600), kinetic tests, etc. |
Nucleic acid analysis (dsDNA, ssDNA, RNA), protein analysis whose absorption varies depending on the concentration at 260 nm or 280 nm (BSA, IgG), cell density (OD600) |
Footprint (W * D) |
216 mm * 290 mm |
145 mm * 190 mm |
Weight |
3.0 kg |
1.4 kg |
Optical specifications |
|
|
Detector |
CMOS linear image sensor (2048 pixels) |
Silicon photodiode |
Light Source |
Xenon fash lamp |
LEDs |
Wavelength range |
190 – 850 nm (full spectrum) |
260 nm, 280 nm / 600 nm (cuvette) / 360 nm (baseline) |
Measurement time |
< 8 seconds |
< 10 seconds |
Absorbance accuracy |
3% (at 0.97A at 302 nm) |
3% (at 1A at 260 nm) |
Processing / Power / Software |
|
|
Software compatibility |
Windows® 7 and 10 |
– |
CPU |
Octa Core ARM® CortexTM-A53 Processor |
– |
Storage |
32 GB internal storage |
8 GB internal storage |
Connectivity |
4 x USB ports, Ethernet, and RS-232 |
2 x USB ports, USB-B, and RS-232 |
Operating systm (OS) |
Android |
Firmware |
Display |
|
|
Display |
7-inch, 1280 x 800 HD color display |
4.3-inch, 480 x 272 color display |
Touchscreen |
Multipoint capacitive touch |
Resistive touch |
Glove compatibility |
Compatible with lab gloves |
Compatible with lab gloves |
Microvolume performance |
|
|
Minimum sample volume |
1 µL |
2 µL |
Photometric range |
0.02 – 330 A (10 mm equivalent) |
0 – 200 Abs |
Detection limit (microvolume) |
2 ng/μL (dsDNA), 0.06 mg/mL (BSA), 0.003 mg/mL (IgG) |
2 ng/μL (dsDNA), 60 mg/mL (BSA), 28,8 mg/mL (IgG) |
Maximum concentration |
16,500 ng/μL (dsDNA), 400 mg/mL (BSA) |
2,000 ng/µL (dsDNA), 60 mg/mL (BSA), 28.8 mg/mL (IgG) |
Cuvette performance |
|
|
Detection limit (cuvette) |
0.2 ng/μL (dsDNA), 0.006 mg/mL (BSA), 0.0003 mg/mL (IgG) |
– |
Photometric range |
0 – 2 A |
0 – 2 A |
Cuvette center height |
8,5 mm |
15 mm |
Only logged in customers may leave a review.